Measuring a microphone’s self-noise without an anechoic chamber

Talas One · workshop notes · 21 November 2026

A studio condenser microphone hisses a little on its own. That hiss, its self-noise, is the number in the datasheet: 5 dB(A) for a RØDE NT1-A. Measuring it at home is harder than it looks, because the room is almost always louder than the microphone. This page is what we learned trying to measure it for the Talas One, an open-hardware dual-diaphragm microphone: what failed, what works, and the protocol for the next attempt.

In short. In a bedroom at night you only measure the room. A dummy capsule gives the noise of the electronics alone: about 5 dB(A) on our channel A, three takes within 0.1 dB. The figure for the complete microphone needs a place quieter than the microphone, and a second microphone as a witness. That measurement is not done yet; this page will carry it when it is.

1Why the room wins

Noise from the room adds to the microphone’s own. The rule given by practitioners is a background 8 to 10 dB below the microphone’s noise (recording.org). An ordinary home sits around 25–40 dB(A). To read 5 dB(A) honestly you would need a room near −5 dB(A): no home gets there.

Two consequences follow, and they organise everything below:

“Anechoic” means without echo, not silent. A semi-anechoic room can still be too loud for this: one university room states a background below 17.5 dB(A). What to ask any lab is its background noise in dB(A).

2What failed: the bedroom at night

On 8 November 2026 we put the Talas One P3 and the NT1-A side by side in a bedroom at night, gains matched within 0.3 dB. Our measurement bench refused the run. Its guard looks at the coherence between the two microphones: if they hear the same thing, they are hearing the room, not themselves. Coherence was 0.20 against a limit of 0.05, in every band from 20 Hz to 20 kHz.

A-weighted, the two microphones read within 0.6 dB of each other, 11 dB above the interface’s own floor. That is still worth something: in the same room, the P3 is no noisier than an NT1-A. But it is not a self-noise figure.

3What works: a dummy capsule

The electronics can be measured without any sound at all. Unsolder one diaphragm’s wire, insulate it, and solder in its place a capacitor of the same value as the diaphragm. The preamp then sees exactly the impedance it sees in use, but nothing acoustic reaches it. Whatever comes out is the noise of the electronics.

Channel A, 44 pF dummy11 November13 November (three takes)
Everything, A-weighted23.2 dB(A)5.1–5.2 dB(A)
Share of 100 Hz–5 kHz power on mains lines92 %12 %
Mains lines removed≈ 13.4 dB(A)5.6–5.7 dB(A)

The first figure was mostly mains hum: 92 % of the power sat on the lines of 50 Hz and its harmonics. Two faults explained it, a body whose parts touched badly and three wrong capacitors on the XLR plugs; they are the subject of our page on hum. Once fixed, the same measurement gave 5.1–5.2 dB(A), repeatable within 0.1 dB, and 3 dB below what our simulation predicted.

What this figure is not: the self-noise of the microphone. A real capsule adds its own noise, and our second channel has not been measured with a dummy yet.

4The complete microphone: the options

MethodWhat it gives
Dummy capsuleThe electronics alone, ±2 dB
Side by side with a reference, at nightAn upper bound
Wardrobe, blanketsA lower upper bound: textiles stop the highs, not the city’s rumble
Underground cellar, box of clothesPossibly a valid figure: earth and rock stop the low end that textiles cannot
Steel pot with a tight lid, in a box of blanketsA good upper bound, possibly a figure
Add a factory-calibrated measurement microphoneRemoves the ±2 dB of a relative sensitivity
Massive steel enclosure, as manufacturers useThe figure
University anechoic chamber with a quiet background, or a testing labThe figure, with a report

A 94 dB acoustic calibrator is often suggested. It does not fit here: it is made to seal on a ½″ or ¼″ measurement microphone, not on a 34 mm capsule behind a grille.

5Our next attempt: a rock wine cellar

A cellar dug in rock, about a metre underground, is the best free place we have: the earth is the mass that stops traffic rumble. Wine cellars, though, run at 70–85 % relative humidity, and a condenser capsule with a 1 GΩ input node does not like moisture: leakage shows as crackles and a rising low end.

The NT1-A (5 dB(A)) readsThe microphone under test readsConclusion
2–8 dB(A)anythingA valid figure
above 8at or below the targetTarget met (an upper bound below it)
above 8above the target, same as the NT1-AInconclusive: the place is too loud
above 8clearly above the NT1-AThe microphone is noisier than the place: check humidity, then the capsule

6The tools

Everything above runs on mic-test-bench, our measurement bench, free software under AGPL-3.0: bruit_fausse_capsule.py for the dummy capsule, campagne_bruit.sh for a side-by-side self-noise run, with the coherence guard that refused our bedroom. The full protocol, with the sheet to fill in on the spot, is in the Talas One repository, and every measurement is in measurements-p3.md.

The Talas One’s circuit is the one Jules Ryckebusch (DJJules) published in 2021 (“True Condenser OPA Mics”, CC BY 4.0). The boards, the measurements and this text are Talas’s work, under CC BY-SA 4.0.